Technical Information:
Table 1 Properties of MediaScout ® MaxiChrom 100-X columns (two examples with different bed heights outlined)
| Column Type | Inner Diameter X Length | Column Cross Section | Column Volume | Flowrate1 | Max. Flowrate1 |
| | (mm x mm) | cm2 | ml | ml/min / cm/h | ml/min / cm/h |
| MaxiChrom 100-127 | 100 x 127 | 78.5 | 1000 | 300 / 229 | < 800 / 611 |
| MaxiChrom 100-200 | 100 x 200 | 78.5 | 1500 | 300 / 229 | < 800 / 611 |
1This depends on the chromatography material
2 The chemical stability refers to the column hardware parts only. The individual packed chromatography media may have different stability. To avoid damage to the chromatography media, please refer to the relevant manufacturer’s data.
|
 |
Resin packing: flow packing to defined compression, according to resin specification |
 |
Connectors:standard Mini TriClamp® ½’’ |
 |
Performance: depending on chromatographic medium |
 |
Pressure maximum: 1.5 bar |
 |
Temperature range: Operation/Storage: 5°C - 60°C / 5°C - 30°C |
 |
Storage conditions: equilibrated with 20% ethanol (ion exchangers with additional 150 mM sodium chloride) |
 |
Materials: Polyoxymethylene (POM) (column tube, filter adaptor, end piece, end cap), Viton® (o-rings) |
 |
Filter plates (frits): Polypropylene/polyethylene (17 µm pore size, others on inquiry) |
 |
Chemical stability2: Columns are tolerant to aqueous buffers and salt solutions, pH 2-14, 1 M alkaline solutions, 1 M non-oxidizing mineral acids, 8 M urea, 6 M guanidine hydrochloride, ethanol, isopropanol and detergents1. They are not compatible with strong oxidants and most organic solvents |
|
|