Atoll
Information

Technical Information:


Table 1 Properties of MediaScout® MaxiChrom 100-X columns (two examples with different bed heights outlined)

Column Type

Inner Diameter X Length

Column Cross Section

Column Volume

Flowrate1

Max. Flowrate1

                                      

(mm x mm)

cm2

ml

ml/min / cm/h

ml/min / cm/h

MaxiChrom 100-127

100 x 127

78.5

1000

300 / 229

< 800 / 611

MaxiChrom 100-200

100 x 200

78.5

1500

300 / 229

< 800 / 611

1This depends on the chromatography material
2 The chemical stability refers to the column hardware parts only. The individual packed chromatography media may have different stability. To avoid damage to the chromatography media, please refer to the relevant manufacturer’s data.



dot

Resin packing: flow packing to defined compression, according to resin specification

dot

Connectors:standard Mini TriClamp® ½’’

dot

Performance: depending on chromatographic medium

dot

Pressure maximum: 1.5 bar

dot

Temperature range: Operation/Storage: 5°C - 60°C / 5°C - 30°C

dot

Storage conditions: equilibrated with 20% ethanol (ion exchangers with additional 150 mM sodium chloride)

dot

Materials: Polyoxymethylene (POM) (column tube, filter adaptor, end piece, end cap), Viton® (o-rings)

dot

Filter plates (frits): Polypropylene/polyethylene (17 µm pore size, others on inquiry)

dot

Chemical stability2: Columns are tolerant to aqueous buffers and salt solutions, pH 2-14, 1 M alkaline solutions, 1 M non-oxidizing mineral acids, 8 M urea, 6 M guanidine hydrochloride, ethanol, isopropanol and detergents1. They are not compatible with strong oxidants and most organic solvents